來源:上海滬升實驗儀器廠發布(bu)時間:2018-03-28
在高溫(wen)(wen)試驗中,如溫(wen)(wen)度變化達不到試驗溫(wen)(wen)度值時,可(ke)以檢查電器系統(tong),逐(zhu)一排除故障(zhang)。如溫(wen)(wen)度升(sheng)得很慢,就要(yao)查看風循(xun)環系統(tong),看一下風循(xun)環的調節擋板(ban)是(shi)否開啟(qi)正常(chang),反之,就檢查風循(xun)環的電機(ji)運轉(zhuan)是(shi)否正常(chang)。如溫(wen)(wen)度過沖厲害那么(me)就需要(yao)整定(ding)PID 的設置參數。如果溫(wen)(wen)度直接上(shang)升(sheng),過溫(wen)(wen)保護,那么(me),控(kong)制器出故障(zhang),須更換控(kong)制儀表。
在做(zuo)濕熱試驗中(zhong)(zhong)(zhong),出(chu)現(xian)實(shi)際濕度(du)會達到100%或(huo)者實(shi)際濕度(du)與目標(biao)濕度(du)相差很大(da),數值低得很多(duo),前(qian)者的(de)現(xian)象:可能是(shi)(shi)(shi)濕球傳(chuan)(chuan)感(gan)器(qi)上的(de)紗布干燥(zao)引起,那就要(yao)檢查濕球傳(chuan)(chuan)感(gan)器(qi)的(de)水(shui)槽中(zhong)(zhong)(zhong)是(shi)(shi)(shi)否缺水(shui),水(shui)槽中(zhong)(zhong)(zhong)的(de)水(shui)位(wei)(wei)是(shi)(shi)(shi)由一水(shui)位(wei)(wei)控(kong)制器(qi)自動控(kong)制的(de),查水(shui)位(wei)(wei)控(kong)制器(qi)供水(shui)系統是(shi)(shi)(shi)否供水(shui)正常(chang),水(shui)位(wei)(wei)控(kong)制器(qi)工(gong)作是(shi)(shi)(shi)否正常(chang)。
另一種可能就是濕球紗布因使用時間長,或供水水質純凈度的原因,會使紗布變硬,使紗布無法吸收水份而干燥,只要更換或清洗紗布即可排除以上現象。后者的現象主要是加濕系統不工作,查看加濕系統的供水系統,供水系統內是否有一定的水量,控制加濕鍋爐水位的水位控制是否正常,加濕鍋爐內的水位是否正常。如以上一切都正常,那就要檢查電器控制系統,這要請專業維修人員進行檢修。
低溫達不到試驗的指標,那你就要觀察溫度的變化,是溫度降的很慢,還是溫度到一定值后溫度有回升的趨勢,前者就要檢查一下,做低溫試驗前是否將工作室烘干,使工作室保持干燥后再將試驗樣品放入工作室內再做試驗, 工作室內的試驗樣品是否放置的過多,使工作室內的風不能充分循環,在排除上述原因后,就要考慮是否是制冷系統中的故障了,這樣就要請廠家的專業人員進行檢修。后者的現象是設備的使用環境不好所致,設備放置的環境溫度,放置的位置(箱體后與墻的距離)要滿足要求(在設備操作使用說明中都有規定)。
設備在試驗運行過程中突然出現故障時,控制儀表上出現對應的故障顯示提示并有聲訊報警提示。操作人員可以對照設備的操作使用中的故障排除一章中快速檢查出屬于哪一類故障,即可請專業人員快速排除故障,以確保試驗的正常進行。其它環境試驗設備在使用中還會有其它的現象,那就要具體現象,具體分析和排除。
上一篇:沖擊試驗箱的簡單維修常識
版權所有(you) 上海滬升儀器設(she)備有(you)限公司
咨詢熱線:400-668-1755
卜先(xian)生:138 1634 3163
傳真: E-mail:
地(di)址(zhi):上海(hai)市奉(feng)賢區青村鎮金(jin)錢公路(lu)4969號 網(wang)站(zhan)設計(ji):